• Confocal Raman Analysis
• Gas Detection
|
ST50S |
ST90S |
ST100S |
detector |
Effective pixels |
512 * 1 |
1340 * 100 |
2000 * 256 |
Cell size |
25 μm * 500 μm |
20 μm * 20 μm |
15 μm * 15 μm |
Fit the camera |
PI, Andor and other scientific research cameras (optional) |
Refrigeration temperature |
-60℃ ~ -80℃ |
Optical parameters |
Wavelength range |
534 nm~665 nm |
790 nm~970 nm |
1080 nm~1330 nm |
Optical resolution |
5 cm-1 (25 μm)8 cm-1(50 μm) |
6 cm-1 (25 μm)8 cm-1(50 μm) |
5 cm-1 (25 μm)8 cm-1(50 μm) |
focal length |
50 mm |
90 mm |
100 mm |
raster |
Body holographic transmission grating |
The width of the incident slit |
5, 10, 25, 50 μm or as customized according to your needs |
Incident optical interface |
SMA905 fiber optic interface, free space |
Electrical parameters |
Integration time |
1 ms ~ 60 min |
Data output interface |
USB 2.0 |
ADC bit depth |
16 bit |
Power supply |
DC11 to 13 V (type @12 V) |
Operating current |
3 A |
Operating temperature |
-20°C~60°C |
Storage temperature |
-30°C~70°C |
Operating humidity |
< 90% RH (non-condensing) |
Physical parameters |
size |
267 * 215 * 109 mm(Detector not included) |
267 * 215 * 109 mm(Detector not included) |
185 * 150 * 79 mm(Detector not included) |
weight |
3.9 kg (without detector) |
4.3 kg (without detector) |
2.2 kg (without detector) |
ST90S transmission imaging spectrometer for testing anhydrous ethanol
(Laser power: 100mW, Exposure time: 5ms)
ST90S transmission imaging spectrometer multi-channel detection